Most clogged column and/or tip problems can often be
easily diagnosed with visual examination using a transmitted light microscope.
The most common reasons that columns become clogged
are as follows:
Contamination caused by water
Using in-house meg-ohm water can cause
carbon particulates to build up within columns. Carbon build-up
will often result in a black band at the head of the column.
Using HPLC-grade bottled water reduces the likelihood of
clogging due to particulate contamination.
Improperly cut fused silica
Improper cleaving of fused-silica tubing
can lead to silica shards in your column or tip. These particles
will be transparent and have sharp, fractured edges. Always
be sure to follow proper cleaving procedures and cleave the
end of all columns and fused-silica tips prior to use. Contaminated
filters and unions should be replaced.
Incorrect self-packing
It is very easy to allow contaminants and
foreign particles to be lodged in your column bed during
the packing process. New Objective offers quality pre-packed
columns, which are all inspected prior to shipment. Customer-supplied
materials can be custom packed.
All New Objective columns and tips are
visually inspected prior to shipping to have an open
bore and to be clean and free of obstruction.