Choosing the Right Tip

STEP 1: Determine if you need an online or offline emitter

Are you doing online nanospray, microspray and LC-MS?

Online analysis is when mobile phase is pumped, at low flow rates, through a small diameter emitter into a mass spectrometer. High voltage is applied to the mobile phase to generate an even spray into the mass spectrometer. The source of the mobile phase, most commonly a syringe pump or a liquid chromatograph (LC), delivers sample that has passed through chromatographic media.

I'm doing online nanospray >>

     Are you doing offline, static nanospray?

Offline analysis is when a discrete sample is loaded by syringe or pipette into an emitter that is then mounted into the mass spectrometer's source. To generate the sample flow offline techniques rely on the electrostatic attraction of the liquid inside the emitter, which has been charged by application of high voltage, towards the counter-electrode.



I'm doing offline nanospray >>

MethodPumped Flow*Sample VolumeTip IDFlow Rate
Offline No0.1 to 5µL
(1 µL is typical)
1 - 4 µm10 to 80 nL/min
(25 µL/min typical)
OnlineYes*<0.1 to >10 µL
(1 - 5 µL is typical)
2 - 50 µm
(5 - 15 µm is typical)
100 nL to - 1 µL/min
(50 - 500 µL/min is typical)
* Pumping may be pressurized or electro-osmotically driven.